Optimize your QC testing while remotely monitoring the process using fiber optic simultaneous multiplexing technology and integrated communication for real-time feedback with the Thermo Scientific™ Antaris™ MX FT-NIR Process Analyzer.

Highlights

  • Thermo Scientific™ ValPro™ Instrument Qualification Package
  • Always-on internal background
  • SMA905 industry-standard connectors for fiber optics — choose the probe right for your process
  • OPC, 4-20mA and Digital I/O for process communications
Brand: Thermo Scientific™

Product Overview

Description

The Antaris MX FT-NIR Process Analyzer provides a complete, fit-for-purpose solution for point-of-use materials analysis and online process monitoring applications.
This easy-to-use NIR analyzer allows a single process to be monitored at multiple points or multiple processes to be monitored simultaneously. The Antaris MX Process Analyzer also eliminates removing the probe from the process stream for background collection or qualification.

Features

The Antaris MX FT-NIR Process Analyzer Features:
  • Thermo Scientific™ ValPro™ Instrument Qualification Package
  • Always-on internal background
  • SMA905 industry-standard connectors for fiber optics — choose the probe right for your process
  • OPC, 4-20mA and Digital I/O for process communications
Recommended for:
  • In-process liquid analysis
  • In-process solid analysis
  • Laboratory liquid analysis
  • Laboratory solid analysis

Specifications

Description Antaris MX FT-NIR Process Analyzer
Analyzer Type Fourier Transform Near-Infrared
Detection High sensitivity, high stability matched InGaAs
Height (English) 18.97 x 15.98 x 9.72 in.
Height (Metric) 48.2 x 40.6 x 24.7 cm
Interface Direct PC to ethernet allows file system and OPC communications
Interferometer Proven, frictionless, stable, long-life Michelson
Operating Temperature Range 15°C to 35°C (59° to 95°F)
Photometric Linearity Slope 1.0 ±0.05 and an intercept of 0.0 ±0.05
Resolution 4 cm-1 across spectral range (0.6 nm at 1250 nm), 2 cm-1 option across spectral range (0.3 nm at 1250 nm)
Sampling Mode Fiber optic multiplexing
Sealed and Desiccated Yes
Source Type Long-life, high-intensity halogen NIR source; spare source included with system, guaranteed filament image alignment
Spectral Range 12,000 to 3800 cmto1 (833 to 2500 nm)
System Status Indicators Indicator lights report scan, laser, power, and source status continuously
Voltage 90/264 V
Wavenumber Accuracy ±0.1 cm-1 (0.02 nm at 1250 nm)
Wavenumber Repeatability (System to System) Better than 0.05 cm-1 (0.008 at 1250 nm)
Wavenumber Reproducibility (Single System) 10 measurement standard deviation <0.006cm-1
Weight (English) 105.16 lb.
Weight (Metric) 25.8 kg